red-y industrial series

คุณสมบัติสินค้า:

Reliable technology and industry standard interfaces for rough environments: Our tried and tested thermal mass flow meters and controllers for gases of the red-y smart series now available as IP67 / NEMA 6 version.

Share

Accurate Measurement
The massflow devices offer high accuracy and a wide dynamic range with the 2 different instrument versions “Standard” and “Hi-Performance”:

Accuracy up to ± 0.3% of full scale + ±0.5% of reading
Turndown ratio 1 : 100 (extended turndown ratio on request)

Analog & Digital: 2 in 1
The red-y industrial mass flow meters and controllers make use of the latest MEMS technology and have a digital (Modbus RTU) and analog interface as standard.

Profibus, Profinet RT and EtherCAT interfaces are available as option.

Software get red-y
Efficient device setup with the free get red-y software:

  • Service tool for remote maintenance
  • Switch gas type
  • Switch measurement units
  • Adjust control parameters

Optional modules get red-y software:

  • Datalogging
  • Gasmixing

Tech Specs

Measuring ranges

Air/Full scale freely selectable

Body size ¼” from 0 … 25 mln/min to 0 … 60 ln/min (from 0 … 27 SCCM to 0 … 64 SLPM)
Body size ½” from 0 … 60 ln/min to 0 … 450 ln/min(from 0 … 65 SLPM to 0 … 480 SLPM)

 

Instrument versions

“Standard”
The economic solution
Accuracy: ± 1.0% of full scale*
Turndown ratio: 1 : 50
“Hi-Performance”
With highest accuracy and turndown ratio
(available for GIM < 200 ln/min (200 SLPM) /
GIC < 150 ln/min (150 SLPM) (air))
Accuracy: ± 0.3% of full scale + ± 0.5% of reading*
Turndown ratio: 1 : 100
*An additional error of ±0.25% may apply for analogue signals

 

Performance Data

Media (real gas calibration)Air, O2*, N2*, He, Ar, CO2, H2, CH4, C3H8 (other gases and gas mixtures on request)
*O2 & N2 are calibrated with air
Response timeMeter (GIM): ± 80ms*; Controller (GIC): ± 500ms*
*depending on device configuration & according to SEMI standard E17-1011,
5-100% of range under optimized conditions
Repeatability± 0.2% of full scale (according to SEMI standard E56-0309)
Longterm stability< 1% of measured value / year
Power supply24 Vdc (18 – 30 Vdc), 15 Vdc on request
Operation pressure0.2 – 11 bara (3 – 160 psia) / GIC with valve type 4.5 and 8: max. 8 bara (120 psia)
Temperature (environment/gas)0 – 50°C (32 – 122°F)
Warm-up time< 1 sec. for full accuracy

 

Materials

Body MaterialsStainless steel 316L (see operating instructions for wetted parts)
Electronic HousingAluminum
SealsEPDM (FDA), optional FKM and FFKM

 

Integration & Installation

Output signals analog0-20 mA, 4-20 mA, 0-5 V, 1-5 V, 0-10 V, 2-10 V
Output signals digitalRS-485; Modbus RTU (Slave); LabView-VIs available / option: Profibus DP-V0, DP-V1
Process connectionG¼” (BSPP* female) up to 60 ln/min (64 SLPM)
G½” (BSPP* female) up to 450 ln/min (480 SLPM)
*British Standard Pipe Parallel
Inlet sectionNone required
Electrical connectionCable gland with compression fitting / Option: M12 plug (DIN-standard)
(both connection IP67 protected)
Mounting orientationAll orientations are possible. We recommend horizontal mounting.
Please contact the manufacturer for further information.

 

Safety

Test pressure16 bara (240 psia)
Leak rate< 1 x 10-6 mbar l/s He
Ingress protection classIP67 (conforms to NEMA 6)
EMCEN 61326-1
ATEX CertificationII 3G nA IIC T4 Gc (Category 3 / Zone 2)
II 3D Ex tc IIIC T100°C Dc (Category 3 / Zone 22)


Powered by MakeWebEasy.com